Vista ™ basic
Quick Reference
Program Mode
1. Press OPTS until PROGRAM MODE is displayed.
2. Press on.
3. Enter the rate for Step 1 and press OK.
4. Enter the volume for Step 1 and press OK.
5. Repeat Steps 3 and 4 until all steps are complete and press OK.
6. Review the Total VTBD and press OK.
7. Press START.
Ramp Taper Mode
1. Press OPTS until RAMP TAPER MODE is displayed.
2. Press ON.
3. Enter the total bag volume to be delivered and press OK.
4. Enter the Ramp Up time in hours and minutes and press OK.
5. Enter the taper Down time in hours and minutes and press OK.
6. Enter the Total Time for the infusion in hours and minutes and press OK.
7. Review the calculated Max Rate and press OK.
8. Press START.
Important Keys
Continuous Mode
1. Enter the rate using the number keys.
2. Press VTBD.
3. Enter the bag volume using the number keys.
4. Press OK.
5. Press START.
Alarm Troubleshooting
Air in Line!
-
Check for air in air-in-line detector.
-
Check that tubing is properly inserted in air-in-line detector.
-
Remove air from line.
Battery Too Low!
-
Plug pump into wall outlet.
Door Open!
-
Close door.
Downstream Occlusion!
-
Open clamps.
-
Check for kinked tubing.
-
Check IV site.
-
Check occlusion pressure setting.
Pump Stopped!
-
Pump is stopped.
-
Either start infusion or press start/stop to extend hold time 3 minutes.
Silence alarms by pressing
Refer to Instructions for Use for complete operating instructions.
Setting the Pump
1. Press (On/Off).
2. Press button on top of door to open.
3. Insert tubing into pump.
-
Push open tube pincher (A)
-
Place tubing in tube pincher (A)
-
Insert tubing along channel (B)
-
Push tubing into Air In Line Detector (C)
4. Close Door.
5. Is this a new therapy?
-
Yes: clears current infusion data.
-
No: begin programming infusion.
6. Prime Set?
-
Yes: follow prompts to prime tubing.
-
No: begin programming infusion.
Piggyback Mode
1.Press OPTS until PIGGYBACK is displayed
2. Press ON.
3. Press END.
4. Press PIGY.
5. Enter Rate/VTBD or VTBD/Time as in Continuous Mode.
6. Press START.
Changing the Rate
1. While infusing, enter the new rate using the number keys.
2. Press OK.
NOTE: If the OK Key is not pressed within 10 seconds of the change, the infusion will continue at the original rate.
Changing the VTBD
1. Stop infusion.
2. Press VTBD
3. Press (C).
4. Enter the new volume to be delivered.
5. Press OK.
6. Press START.