top of page

Vista ™ basic

Quick Reference

Screen Shot 2021-01-22 at 4.07.42 PM.png

Program Mode

​

1. Press OPTS until PROGRAM MODE is displayed.

​

2. Press on.

​

3. Enter the rate for Step 1 and press OK.

​

4. Enter the volume for Step 1 and press OK.

​

5. Repeat Steps 3 and 4 until all steps are complete and press OK.

​

6. Review the Total VTBD and press OK.

​

7. Press START.

​

Ramp Taper Mode

​

1. Press OPTS until RAMP TAPER MODE is displayed.

​

2. Press ON.

​

3. Enter the total bag volume to be delivered and press OK.

​

4. Enter the Ramp Up time in hours and minutes and press OK.

​

5. Enter the taper Down time in hours and minutes and press OK.

​

6. Enter the Total Time for the infusion in hours and minutes and press OK.

​

7. Review the calculated Max Rate and press OK.

​

8. Press START.

Important Keys

Screen Shot 2021-01-23 at 10.45.53 AM.pn

Continuous Mode​

​

1. Enter the rate using the number keys.

​

​

​

​

​

​

​

​

2. Press VTBD.

​

3. Enter the bag volume using the number keys.

​

4. Press OK.

​

​

​

​

​

​

​

​

5. Press START.

Screen Shot 2021-01-23 at 11.01.16 AM.pn
Screen Shot 2021-01-23 at 11.02.44 AM.pn
Screen Shot 2021-01-23 at 11.04.17 AM.pn

Alarm Troubleshooting

​

Air in Line!

  • Check for air in air-in-line detector.

  • Check that tubing is properly inserted in air-in-line detector.

  • Remove air from line.

​

Battery Too Low!

  • Plug pump into wall outlet.

​

Door Open!

  • Close door.

​

Downstream Occlusion!

  • Open clamps.

  • Check for kinked tubing.

  • Check IV site.

  • Check occlusion pressure setting.

​

Pump Stopped!

  • Pump is stopped.

  • Either start infusion or press start/stop to extend hold time 3 minutes.

​

Silence alarms by pressing     

​

Refer to Instructions for Use for complete operating instructions.

Screen Shot 2021-01-22 at 4.20.11 PM.png

Setting the Pump

​

1. Press         (On/Off).

​

2. Press button on top of door to open. 

​

3. Insert tubing into pump.

  • Push open tube pincher (A)

  • Place tubing in tube pincher (A)

  • Insert tubing along channel (B)

  • Push tubing into Air In Line Detector (C)

​

​

​

​

​

​

​

​

​

​

​

​​

4. Close Door.

​

5. Is this a new therapy?

  • Yes: clears current infusion data.

  • No: begin programming infusion.

​

6. Prime Set?

  • Yes: follow prompts to prime tubing.

  • No: begin programming infusion.

​

Screen Shot 2021-01-23 at 10.48.53 AM.pn
Screen Shot 2021-01-23 at 10.53.12 AM.pn

Piggyback Mode

 

​1.Press OPTS until PIGGYBACK is displayed

​

2. Press ON.

​

3. Press END.

​

4. Press PIGY.

​

5. Enter Rate/VTBD or VTBD/Time as in Continuous Mode.

​

6. Press START.

Changing the Rate

​

1. While infusing, enter the new rate using the number keys.

​

2. Press OK.

​

NOTE: If the OK Key is not pressed within 10 seconds of the change, the infusion will continue at the original rate.

Screen Shot 2021-01-23 at 11.11.40 AM.pn

Changing the VTBD

​

1. Stop infusion.

​

2. Press VTBD

​

3. Press (C).

​

4. Enter the new volume to be delivered.

​

5. Press OK.

​

6. Press START.

Screen Shot 2021-01-23 at 11.15.55 AM.pn

B. Braun Vista Basic Quick Reference